ISAKOS: 2023 Congress in Boston, MA USA

2023 ISAKOS Biennial Congress ePoster

 

Precision Of Radiostereometric Analysis For Use In Biomechanical Experiments – Evaluation And Comparison With A 3D Camera System

Adrian Deichsel, MD, Münster, NRW GERMANY
Jens Wermers, MD, Münster GERMANY
Michael J. Raschke, MD, Prof., Münster GERMANY
Matthias Klimek, M. Sc., Muenster GERMANY
Christian Peez, MD, Münster, NRW GERMANY
Thorben Briese, MD, Münster GERMANY
Johannes Glasbrenner, MD, Münster GERMANY
Elmar Herbst, MD, PhD, Muenster GERMANY
Christoph Kittl, MD, MD(res), Muenster GERMANY

Department of Trauma, Hand and Reconstructive Surgery, University Hospital Münster, Münster, NRW, GERMANY

FDA Status Cleared

Summary

In this study, Radiostereometric Analysis (RSA) was compared to a commercial 3D camera system for its precision of tracking joint movements in a robotic testing system.

ePosters will be available shortly before Congress

Abstract

Introduction

Radiostereometric Analysis (RSA) is a technique using the concept of digital image correlation (DIC) to determine the position of radiopaque markers in a three-dimensional space. The advantage of RSA over traditional camera-based tracking systems is that no direct optical visibility of the tracked object is necessary, thereby allowing assessment of movements of intraarticular structures. The purpose of this study was to compare a novel RSA setup to a commercial 3D camera system regarding its precision.

Methods

Optical markers (0.4 mm diameter) and tantalum beads (1 mm diameter) were applied to a sawbone knee. The tibia was mounted to the table, the femur was fixed into a six degrees of freedom robotic test setup (KR125; KUKA Robotics), which was optimized to simulate joint movements by using simVITRO (Cleveland Clinic BioRobotics Lab). The robot system has a positional accuracy of 0.05 mm. A protocol consisting of 3 mm of anterior tibial translation (ATT3), followed by 3 mm of medial tibial translation (MTT3), followed by return to the home position, was performed. The protocol was repeated for a total of 10 times. In each step, the position of the femur in respect to the tibia was captured by either a commercial 3D camera system (ARAMIS, GOM GmbH), or the RSA system with one x-ray source and one digital x-ray detector. The changes between positions were calculated and the two systems compared using two-tailed t-est. Correlation between the techniques was performed using Pearson’s correlation.

Results

For ATT3 the measured translation was 3.040 mm (SD 0.005) in the ARAMIS group and 3.039 (SD 0.094) in the RSA group. No significant differences were found between the groups. For MTT3 the measured translation was 2.857 mm (SD 0.020) in the ARAMIS group and 2.798 (SD 0,153) in the RSA group. No significant differences were found between the groups. Correlation between the techniques revealed a near-perfect correlation of 0.97 (95 % CI 0.93 – 0.98).

Conclusion

The presented RSA system can precisely determine motions with an accuracy of < 0.2 mm and is therefore well suited for tracking of joint movements in biomechanical experiments.