Comparison of Range of Motion between Two-Year Clinical Outcomes and Predictions of a Static Scapula Preoperative Planning Software for Reverse Shoulder Arthroplasty
Marco Branni, PhD, Brisbane, QLD AUSTRALIA
Asma Salhi, PhD, Brisbane, QLD AUSTRALIA
Kristine R. Italia, MD, FPOA, Quezon City, Metro Manila PHILIPPINES
Luke Gilliland, BEng, Brisbane, QLD AUSTRALIA
Marine Launay, MEng, Greenslopes, QLD AUSTRALIA
Roberto Pareyon MEXICO
Jashint Maharaj, MBBS, FRSPH, Brisbane, QLD AUSTRALIA
Angus Lane, BEng, Brisbane, QLD AUSTRALIA
Helen Ingoe, MBBS, FRCS (Tr+Orth), MD, MSc, PGCert, Dunedin NEW ZEALAND
Peter Pivonka, PhD, Brisbane, QLD AUSTRALIA
Kenneth Cutbush, MBBS, FRACS, FAOrthA, Spring Hill, QLD AUSTRALIA
Ashish Gupta, MBBS, MSc, FRACS, FAORTHOA, Brisbane, QLD AUSTRALIA
Queensland Unit for Advanced Shoulder Research (QUASR), Brisbane, QLD, AUSTRALIA
FDA Status Not Applicable
Summary
This study aims to compare the range of motion based on preoperative planning software by using the implant position from postoperative CT images (P-ROM), with the clinical range of motion (C-ROM) assessed at minimum of 2 years follow-up.
ePosters will be available shortly before Congress
Abstract
Introduction
Preoperative planning using commercially available software has gained popularity for reverse shoulder arthroplasty (RSA). Three-dimensional automatic segmentation of scapula and humerus offers a visual insight of the pathology, as well providing arc of motion for the implanted articulation and identifying potential areas of bony impingement. However, these software algorithms utilize a fixed scapular model, disregarding the preoperative clinical range of motion of the patient, be it glenohumeral or scapulothoracic, as well as any soft tissue parameters. This study aims to compare the range of motion based on preoperative planning software by using the implant position from postoperative CT images (P-ROM), with the clinical range of motion (C-ROM) assessed at minimum of 2 years follow-up.
Methods
Pre- and post-operative CT-scans of 50 patients who underwent a primary RSA between 2017 and 2021 were analyzed. At the postoperative two-year review, each patient was assessed for active range of motion. Implant size and position based on operative notes and postoperative CT scans were used to replicate the performed surgery into the planning platform. Abduction, flexion, and external rotation motions were simulated and recorded. The relationship between C-ROM and P-ROM was investigated using linear regression analysis, Pearson correlation coefficient, and paired t-test.
Results
P-ROM was significantly lower than C-ROM at two years postoperatively (P<0.001), with an average discrepancy of 77 degrees in abduction, 50 degrees in flexion, and 39 degrees in external rotation (C-ROM: abduction 154±21 (80-180 degrees); flexion 160±17 (90-180 degrees); external rotation 52±14 (10-80 degrees) versus P-ROM: abduction 77±14 (53-107 degrees); flexion 110±24 (67-180 degrees); external rotation 13±19 (0-79 degrees). The linear regression analysis indicated weak agreement between C-ROM and P-ROM (abduction R2=0.01; flexion R2=0.01; external rotation R2=0.07). Pearson’s correlation coefficients revealed weak correlations of -0.09, 0.11 and 0.27 for abduction, flexion, and external rotation.
Conclusion
P-ROM-based preoperative software, in its current form, does not allow the prediction of the C-ROM at two-year follow-up for patients who underwent an RSA. Further research and development are needed to make the ROM simulation feature a more clinically reliable tool that can guide the surgical decision-making.